99mTc MAGS & 99mTc DTPA

COMPARISON TO EVALUATE RENAL FUNCTIONS IN RENAL TRANSPLANT PATIENTS

Authors

  • MUHAMMAD SHAHZAD AFZAL Punjab Institute of Nuclear Medicine Faisalabad
  • Muhammad Aleem Khan Punjab Institute of Nuclear Medicine Faisalabad.
  • NASIR MAHMOOD Centre of Nuclear Medicine, Mayo Hospital Lahore
  • Javaid Irfanullah Punjab Institute of Nuclear Medicine Faisalabad.
  • MUHAMMAD BABAR IMRAN Punjab Institute of Nuclear Medicine Faisalabad.
  • Khan Muhammad Punjab Institute of Nuclear Medicine Faisalabad.
  • Muhammad Saeed Akhtar Punjab Institute of Nuclear Medicine Faisalabad
  • Saeed Khan Mayo Hospital Lahore

DOI:

https://doi.org/10.29309/TPMJ/2003.10.02.5294

Keywords:

Renal functions, renal transplant, DTPA, MAGS, Perfusion Index, Parenchymal retention Index

Abstract

Background: Renal scintigraphy evaluates the renal function non-invasively. The present study was conducted to compare 99mTc MAG3 and
99mTc DTPA in renal transplant patients for the evaluation of renal functions. Materials & Methods: 99mTc MAG3 and 99mTo DTPA studies
were performed on 33 renal transplant patients and 10 normal subjects, with 1 day interval in between the two studies. Renal transplant patients
were categorized as normal, acute tubular necrosis, rejection, urine outflow obstruction and urinary tract infection groups on the basis of routine
clinical investigations and preliminary results of scintigraphic studies. Results: Both radiopharmaceuticals, 99mTc MAG3 and 99mTc DTPA
were having equal diagnostic efficacy, however image quality with 99mTc MAG3 was superior to that of 99mTc DTPA. Renograms clearly
differentiated various renal groups. Correlation between glomerular filtration rate (GFR) and effective renal plasma flow (ERPF) was good
(r=0.89 at P<0.00001). In terms of perfusion index (PI) our results with both agents were similar (r=0.96 at p<0.00001), clearly differentiating
all cases of rejection from acute tubular necrosis. PI was raised in rejection cases, while normal in all other groups. In cases of acute tubular
necrosis and rejection, 99mTc MAG3 based parenchymal retention index (R-20/3) was markedly raised, compared to that of 99mTc DTPA
(p<0.00001). Conclusion: 99mTc- MAG3 and 99mTc DTPA have equal diagnostic efficacy in renal transplant patients. However visual
delineations of the kidneys is better with 99mTc MAG3, especially in cases with compromised renal functions.